7 Repair Procedures
7-18
05/04 1006-0453-000
7.9 Gas inlet valve
1. Tube guide plate
2. GIV solenoid switch
3. Captive mounting screws (2)
4. GIV
5. Inlet filter housing
6. Manifold plate
Figure 7-16 • Gas Inlet Valve (GIV)
Follow the instructions in Section 7.6 to remove the Vent Engine.
To remove the solenoid switch:
1. Use a 1.5 mm hex wrench to remove the mounting screws from the GIV
solenoid switch on the inlet valve.
To replace the gas inlet valve:
1. Remove the tube guide plate from the top of the GIV.
2. Remove the inlet filter underneath the manifold plate.
3. Use a 3 mm hex wrench to loosen the two captive inlet valve mounting
screws.
4. Remove the GIV/inlet filter housing assembly.
5. To separate the GIV from the inlet filter housing, remove the two mounting
screws from within the filter housing.
6. Examine the valve seat and O-rings in the manifold for damage. Replace as
necessary.
7. Reassemble in reverse order.
8. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.