2 Theory of Operation
1006-0453-000 05/04
2-11
2.4.4 Power supply
(integrated CPU)
Aestiva machines with an integrated CPU board use a universal power supply
for AC to DC conversion. The remainder of the power supply functions are
derived in the power supply circuits on the integrated CPU:
• DC to DC step-down converter
• Battery charger
• Multiple output DC regulator
• Battery charge/discharge current monitor
• Battery voltage monitor
• Task light power supply 12V
Figure 2-10 • Ventilator power supply (Integrated CPU)
AB.29.081
-15V ANALOG
SUPPLY LDO
POWER CORD
AC INLET
AC TO DC POWER SUPPLY
85 -- 264 VAC
VCHGR = 16.0 TYPICAL
VMAIN = 12.5V TYPICAL
MOSFET
SWITCHES
SELECT VMAIN OR VBAT
AS OUTPUT TO VBUS
DIODE OR
VMAIN
VCHGR
VBAT
VBUZZER
OUTPUT TO VDD FAIL
BUZZER
VSWITCH
ON / STANDBY SWITCH CIRCUIT
12V BATTERY
THERMAL AUTO
RESETABLE FUSE
BATTERY CHARGER
BULK CHARGE = 400mA
CHARGER and TRICKLE
DISABLE
VOLTAGE and CURRENT
MONITOR
VBUS
TASK LIGHTS LDO
TYPICALLY 11.718V
VHEL DISPLAY LDO
TYPICALLY 11.718V
SIMPLE SWITCHER
VALVES SUPPLY
5.8V OUTPUT and 10VA
VDD DIGITAL
OUTPUT LDO
5V +/- 2.5%
DUAL OUTPUT
SWITCHER
POSITIVE OUTPUT
NEGATIVE OUTPUT
SIB_12V OUTPUT
and 10VA LDO
+15V ANALOG
SUPPLY LDO
VBatt, IBatt
Rem_On
AC_LED
AAAA
BBBB
....2222
9999
....0000
8888
1111
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......