2 Theory of Operation
2-12
05/04 1006-0453-000
2.4.5 Sealed lead acid
battery
A sealed lead acid battery supplies battery backup for the Aestiva 7900
Ventilator. The Aestiva 7900 Ventilator is not a portable unit. Batteries for the
ventilator are used as back up power in case of a power failure. Thus the
battery is in a float charge state most of the time. Batteries meet the
following:
• Capacity to operate unit for 30 minutes.
• Long float charge life.
• Battery pack is internally fused - in line replaceable.
• Battery terminals and connecting wires are protected against short circuits.
IIIInnn
nppp
puuu
utttt
Nominally 13.7 VDC at 25
o
C during float charge.
O
O
O
Ouuu
uttttppp
puuu
utttt
+10V to +14.8VDC during discharge