3 Post-Service Checkout
3-2
05/04 1006-0453-000
3.1 Post-service checkout
After servicing the Aestiva 7900 Ventilator, run the service menu tests that
are pertinent to the components replaced. Perform calibration on flow
sensors, pressure sensitivity, flow valve and bleed resistor.
Then you must complete the checkout procedure for the entire machine:
• the Aestiva 7900 Ventilator,
• the Aestiva Anesthesia Machine,
• and all the accessories and options.
w
WARNING
You must perform all post-service checks after maintenance or service of
the ventilator. Failure to do so may result in patient injury.
w
WARNING
All components and accessories must be connect correctly. All hoses
and cables must be properly connected before returning the anesthesia
machine to clinical use. Failure to do so may result in patient injury.
3.1.1 Test the Aestiva
7900 Ventilator
Perform the Preoperative Checkout Procedure in Part 1 of the Aestiva
Operation Manual.
3.1.2 Test the Aestiva
Anesthesia Machine
The Aestiva 7900 Ventilator is an integral part of the complete Aestiva
Anesthesia Machine. To be certain the ventilator is functioning correctly, test
the entire system. Refer to the Aestiva Anesthesia Machine Service Manual
for the proper checkout procedures.
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......