4a Tests and Calibration 4.X Software
4a-16
05/04 1006-0453-000
4a.3.7 Test GIV
To test the GIV (gas inlet valve) the software first closes the GIV. It reads the
A/D channel for the GIV. If the A/D channel for the GIV does not read closed,
the test fails; otherwise, the test continues.
The software then opens the flow valve to the calibrated flow of the bleed
resistor and waits for the flow to stabilize (about 2 seconds). Once the flow is
stabilized the software checks to see if manifold pressure has dropped to less
than 0.5 cm H
2
O. If manifold pressure does not drop below 0.5 cm H
2
O, the
test fails; otherwise, the test continues.
The software then closes the flow valve and opens the GIV. It reads the A/D
channel for the GIV. If the A/D channel for the GIV does not read open, the test
fails.
Figure 4-17 • Test GIV menu
GIV Test FAIL instructions
Failure can be caused by the valve or the CPU board malfunctioning.
Use a multimeter to measure the resistance of the GIV solenoid. It should be
approximately 25 ohms. If necessary, replace the GIV solenoid
(see section 7).
Use a multimeter to measure the voltage at the inlet valve connector on the
pneumatic connection board. The voltage should be 4.5 volts when the GIV is
open.
• If the voltage is 4.5 volts or greater, service the GIV (see section 7).
• If the voltage is less than 4.5 volts replace the CPU board (see section 7).
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......