4a Tests and Calibration 4.X Software
1006-0453-000 05/04
4a-23
Notes:
1. These columns show the acceptable range where possible.
2. The A/D count and displayed value shown for Airway Pressure, Manifold Pressure, Inspiratory Flow, and Expiratory Flow are the zero
pressure values. These readings should be taken with the flow sensors disconnected by removing the circuit module.
3. The count range and displayed value shown for GIV Current Sense is with the GIV turned on. If the GIV is turned off, the off count range
is 0 to 259 Counts.
Table 4a-1 A/D Channels
A/D Channel
Counts
1
Actual Range
1
Displayed Range
Special Instructions
Inspiratory Flow
2
1800-2300
Near Zero L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050
Counts)
Expiratory Flow
2
1800-2300
Near Zero L/min
-120 to 120 L/min
Zero Offset Reading (nominal 2050
Counts)
Airway Pressure
2
550-1050
Near Zero
cm H
2
O
-20 to 120
cm H
2
O
Zero Offset Reading (nominal 800 Counts)
Manifold Pressure
2
550-1050
Near Zero
cm H
2
O
-20 to 120
cm H
2
O
Zero Offset Reading (nominal 800 Counts)
O2
0-4095
0 to 100%
0 to 100%
Count weight and limits are determined
during O2 calibration
Flow Current Sense
0-4095
0 to 1.024 Amp
0 to 1.024 Amps
Flow DAC Feedback
0-4095
0 to 4.095 Volts
0 to 4.095 Volts
Battery Voltage
1740-3976
7 to 16 Volts
0 to 16.48 Volts
<7V = Low battery voltage failure.
<11.65V = Low battery voltage warning.
>16 Volts (10 Sec) = High battery voltage
failure.
Battery Current
700-4000
-600 mA to 6 Amps
-2 to 6.19 Amps
-150 to -600 mA = Battery charging.
-601 mA to -2 Amps = Charge current out of
range.
>300 mA = Operating on battery.
>6 Amps = Battery discharge current out of
range.
1.225 Voltage
Reference
1214-1235
1.214 to 1.235 Volts
0 to 4.095 Volts
5.8V Voltage Test
1997-2565
4.86 to 6.24 Volts
0 to 9.96 Volts
+15V Analog Supply
1914-2262
13.77 to 16.27 Volts
0 to 29.46 Volts
-15V Analog Supply
858-953
-12.62 to -17.46 Volts -61 to 1.43 Volts
+12V SIB Supply
1779-2322
10.44 to 13.62 Volts
0 to 24 Volts
VEL Supply
1645-2675
9.56 to 15.69 Volts
0 to 24 Volts
VSW Supply
1269-2368
9.13 to 17.04 Volts
0 to 29.46 Volts
+12V Light Supply
1879-2217
11.02 to 13 Volts
0 to 24 Volts
In Rev 3.4 software,
Range = 10.00 to 12.73 Volts
GIV Current Sense
3
2371-3884
143.7 to 235.4 mA
0 to 248.2 mA
Off state reading is 0 to 259 counts
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......