4a Tests and Calibration 4.X Software
1006-0453-000 05/04
4a-27
4a.3.17 Flow Valve Test
Tool
The Flow Valve Test Tool is available for test and troubleshooting purposes
only. It allows you to manually control the flow valve setting from 0 (closed) to
120 LPM, in 1 LPM increments, and observe key pressure and flow
measurements on the same screen.
This is mainly used to test the drive gas circuit and MOPV valves as detailed in
the MOPV test procedure (see section 5.4). However, it can also be used for
other troubleshooting procedures whenever a set flow is required.
Figure 4-28 • Flow Valve Test Tool menu
N
N
N
Nooo
otttteeee:::: The flow does not turn off automatically. The flow must be set to zero to
stop flow. Flow will originally be set to zero upon entering this screen.