4b Tests and Calibration 1.X/3.X Software
4b-4
05/04 1006-0453-000
4b.2 Service Mode Confirmation menu
The service calibration mode tests and/or calibrates hardware necessary to
prepare a ventilator in manufacture for factory shipment and to service a
ventilator in the field.
There are two ways to enter the service mode:
• If the machine is turned off, push and hold in the adjustment knob while
setting the system switch to On. Hold the adjustment knob pushed in
until the “Service Mode Confirmation” menu appears. Use the
adjustment knob to highlight “Service Mode”, then push the adjustment
knob to confirm the selection.
• If the machine is already in normal operation, set the Bag/Vent switch to
Bag. Then, press the V
T /
Pinsp, the PEEP, and the menu switches at the
same time to reset the software (powerup). Push and hold the
adjustment knob until the “Service Mode Confirmation” menu appears.
All data displayed by the ventilator during and after a service mode test or
measurement is also output to the serial port.
Figure 4b-1 • Service confirmation menu
From this menu you can:
• Set Altitude
• Set Language
• Set Serial Connection
• Go to the Service Mode or Normal Operation
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......