4b Tests and Calibration 1.X/3.X Software
1006-0453-000 05/04
4b-9
4b.4.5 Test EEPROM
The software tests all of the EEPROM memory via a bit pattern test. It writes a
certain pattern to a block of memory and then reads that block of memory. If
the bit pattern that was written is not the same as the bit pattern read back,
the test fails.
EEPROM test FAIL instructions
Upon FAIL, replace the CPU board (see section 7.5).
4b.4.6 Test Panel
Switches
The software sets up the ventilator to receive all button presses and control
wheel turns. The test asks the user to press each of the buttons and turn the
control wheel one full turn in both directions. Each time a button is pressed
the number next to the button is reverse-highlighted. If the button is pressed
and the number next to the button does not reverse highlight the test for that
button failed. If all of the buttons are pressed and all of the numbers next to
the buttons are reverse-highlighted the button press test passes. The control
wheel turn test works the same way. The user is prompted to turn the control
wheel clockwise and counter-clockwise one full turn (16 clicks). When the
control wheel is turned (one full turn in either direction) and the control wheel
is working, the test passes. If the control wheel is not working, the test times-
out in about 10 seconds and the test fails.
Panel switch test FAIL
instructions
If any of the select buttons or the adjustment knob turn test fails, replace the
front panel keyboard (see Section 7.2) or the rotary encoder (see Section 7.3).
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......