4b Tests and Calibration 1.X/3.X Software
4b-14
05/04 1006-0453-000
4b.4.11 Test 5V Fail
Alarm
This test is only available in 3.x software.
A 5-Volt supply (VDD) is used to power the digital circuits throughout the
ventilator.
Where the 5-Volt supply (VDD) is derived depends on the type of CPU board
that is used:
• In machines with the original CPU board (non-integrated), the VDD voltage is
derived on the power supply board.
• In machines with an integrated CPU board, the VDD voltage is derived in the
power section of the CPU board.
If the VDD supply fails, the ventilator will sound a continuous alarm tone when
the system switch is turned on. The alarm is generated on the CPU board
through a dedicated “sound source” that is part of the CPU board.
To test the 5V Fail Alarm, follow the directions on the screen.
Software 3.X
Remarks
If the alarm tone does not sound, replace the CPU board (see Section 7.5).