4b Tests and Calibration 1.X/3.X Software
1006-0453-000 05/04
4b-17
4b.5.2 Display I/O
signals
The
Discrete I/O Signals
menu displays discrete binary signals associated with
machine switch positions.
There are several types of switches in the Aestiva machine:
•
some switches are mechanically operated,
•
some switches are pneumatically operated,
•
some switches are electronic,
Mechanical switches
• Control Panel Cover — Closed or Open
• CO
2
Bypass — Off or On
• ACGO Switch Status — Ventilation or Aux CGO (machines with ACGO)
• Bag/Vent Status — Bag or Vent
• Canister Status — Closed or Open
Pneumatic switches
• O
2
Supply Pressure Status — Pressure or No Pressure
• O
2
Flush Status — Off or On
• DPL Switch Status — Off or On
Electronic switches
• O
2
Sensor Status — Connected or Disconnected
• Circuit module ID
Off = tab present to block light
On = no tab to block light
Breathing Circuit
Switch
Circle
Bain/Mapleson D
(no module)
One
Off
On
On
Two
Off
Off
On
Three
On
Off
On
Four
On
On
On