4b Tests and Calibration 1.X/3.X Software
1006-0453-000 05/04
4b-21
4b.7 Test Breathing System For Leak
The software allows you to test the breathing system for leaks. By using the
patient circuit to establish a closed loop, you can measure the leak rate. The
leak rate is equal to the fresh gas flow needed to maintain 30 cm H
2
O. The
system should have a leak rate ≤ 200 mL.
Software 3.X
Software 1.X
Remark
Refer to the 3.X software menu to clarify two items on the 1.X software menu:
• In step 2, set the fresh gas flow to minimum.
• In step 6, you can occlude the patient circuit by connecting the open end of
the patient “Y” to the plug that is part of the handle at the front of the
breathing system.