4b Tests and Calibration 1.X/3.X Software
4b-30
05/04 1006-0453-000
4b.12 User Settings
4b.12.1 Select Altitude
This provides the user with the means to select the altitude setting.
4b.12.2 Select Drive Gas
This provides the user with the means to tell the ventilator which drive gas is
being used.
w
Caution
If you change the drive gas, you must also change the drive gas selection
on the service setup screen. If the drive gas selection and the actual
drive gas do not agree, volumes will not be correct.
To change the actual drive gas, refer to Section 4, “Repair Procedures,” of the
Aestiva Anesthesia Machine Service Manual.