5 Troubleshooting
5-26
05/04 1006-0453-000
5.5 Troubleshooting Flowcharts
5.5.1 Ventilator assessment process
* Refer to Section 7.2,
“Breathing System Leak Tests,”
in the Aestiva Machine service manual, 1006-0452-000.
Start
Turn on System
Review Active Alarms,
Error Log and Alarm Log
No
Proceed
to “No Display”
Troubleshooting
No
Use Alarm
Troubleshooting
Proceed to
“Breathing System Leak”
Troubleshooting
Yes
No
Yes
Is Display
w orking?
Diagnosis
achieved?
Done
Alarms of Type:
Patient Circuit Leak
System Leak
Bellows falls or leak
detected
ALARMS
SYMPTOMS
Yes
Yes
Proceed to
“Inaccurate Volume
Ventilation”
Troubleshooting
Yes
Alarms of Type:
Vt Compensation Off
Delivered Volume Mismatch
Check Flow Sensors
Insp Reverse Flow
Exp Reverse Flow
System Leak
Unable to drive Bellows
Inaccurate ventilation
Yes
No
Proceed to
“No Ventilation”
Troubleshooting
Yes
Alarms of Type:
Low Paw
Volume Apnea
Low Drive Gas
Pressure System
Leak
Will not
ventilate
Yes
No
Proceed to
“High Intrinsic PEEP”
Troubleshooting
Yes
Alarms of Type:
Sustained Paw
High Paw
Continuously high PEEP
during operation
Yes
No
No
No
Yes
*
Are
there
alarms?
No
Section 5
AB.29.092
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......