5 Troubleshooting
1006-0453-000 05/04
5-29
5.5.4 No ventilation troubleshooting
* Refer to Section 7.2,
“Breathing System Leak Tests,”
in the Aestiva Machine service manual, 1006-0452-000.
*
*
Start
Calibrate Flow
Sensors
Verify pass?
Proceed to
"Inaccurate Volume Ventilation"
Troubleshooting
No
Yes
Perform All
Calibrations and
Re-evaluate
Does bellows
move?
Yes
Check the output
of the Drive Gas Regulator
Regulator
pressure OK at
25 psi?
Set Flow Valve to
120 L/min
No
Yes
High flow gas
exhausts from
drive gas port?
Test Flow Valve
No
Proceed to
"Breathing System Leak"
Troubleshooting
No
Does bellows
return to top of
canister?
Yes
Check Hospital
pressure supply
No
Hospital
pressure OK?
Verify operation of GIV.
Check/Replace
inlet filter
if necessary.
Yes
Correct Hospital
supply problem
No
Examine Exhalation Valve
and diaphragm
Yes
Does circuit
pressurize?
Proceed to
"Breathing Circuit Leak"
Troubleshooting
No
Check function
of drive pressure
pop-off valve
Yes
AB.29.096
Section 4
Section 4
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......