7 Repair Procedures
7-12
05/04 1006-0453-000
7.5.5 Toroid
(original CPU only)
To remove the toroid, you must first remove the AC inlet module and the
electronics compartment cover. The toroid is located in the electronics
enclosure next to the power supply board.
Figure 7-10 • Isolation transformer (top view)
1. Remove the AC inlet module and the electrical enclosure cover
2. Disconnect the battery cable from the power supply board by pressing the
lock tabs on either side of the connector and gently pulling on the
connector.
3. Disconnect the grounding lug using a 7-mm socket wrench.
4. Remove the two mounting nuts using a 7-mm socket wrench.
5. Remove the toroid.
6. Install a new toroid by following these instructions in reverse order and
reconnecting like-colored connectors.
7. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.
3
1
2
3
3
1. Toroid
2. Grounding lug
3. Mounting nuts