7 Repair Procedures
1006-0453-000 05/04
7-13
7.5.6 Battery
To remove the battery, you must first remove the AC inlet module and the
electronics compartment cover.
Note
Aestiva machines with the original CPU or with the Integrated CPU use the
same battery. However, they are mounted differently.
• The battery for the original CPU is mounted on the back plate of the electrical
enclosure. The enclosure cover includes a foam insert to keep the battery in
place.
• The battery for the Integrated CPU is mounted on the floor of the electrical
enclosure and does not require the foam insert on the enclosure cover.
Figure 7-11 • Battery (original CPU shown)
1. Remove the AC inlet module and the electrical enclosure cover
2. Disconnect the battery cable by pressing the lock tabs on either side of the
connector and gently pulling on the connector.
3. Remove the battery.
• For original CPU machines, the battery is held in place with a metal strap
(and the foam insert). Slide the battery to the right until it clears the strap
and remove it.
• For Integrated CPU machines, the battery is held in place with a formed
bracket and the wall of the electrical enclosure. Remove the bracket to
replace the battery.
4. Install a new battery by following these instructions in reverse order.
5. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.
6. Allow the battery to charge.
2
3
1. Battery cable
2. Battery
3. Battery retainer foam insert
1
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......