7 Repair Procedures
7-16
05/04 1006-0453-000
7.7 Non-relieving regulator
The regulator is replaced as an assembly.
1. Non-relieving regulator
2. Phillips head screws
3. Captive regulator mounting screws
Figure 7-14 • Non-relieving regulator
To remove the non-relieving regulator:
1. Follow the instructions in Section 7.6 to remove the Vent Engine.
2. Do not remove the Phillips head screws (4) that hold the regulator together.
Use a 3 mm hex wrench to loosen the two captive mounting screws.
3. Lift the regulator from the Vent Engine manifold.
4. Inspect seat and O-rings for damage. Replace as necessary before
replacing the regulator.
5. Replace the regulator by following these instructions in reverse order.
6. Perform the regulator calibration procedure in Section 4.
7. Install the Vent Engine by following the steps in Section 7.6 in reverse order.
8. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.