7 Repair Procedures
1006-0453-000 05/04
7-17
7.8 Flow control valve
1. Flow control valve cable
2. Flow control valve
3. Captive mounting screws (2)
Figure 7-15 • Flow control valve
To service/replace the flow control valve:
1. Follow the instructions in Section 7.6 to remove the Vent Engine.
2. Use a 3 mm hex wrench to loosen the two captive mounting screws and lift
out the flow control valve.
3. Examine the seat and two O-rings for damage.
Replace as necessary.
4. Reinstall the flow control valve by following these instructions in reverse
order.
5. Perform the flow valve test in Section 4.
6. Install the pneumatic assembly by following the steps in Section 7.6 in
reverse order.
7. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.