8 Illustrated Parts
8-2
05/04 1006-0453-000
8.1 Special instructions
Apply a thin coat of oxygen-use-approved lubricant to o-rings prior to
installation (unless otherwise noted). Use:
• KRYTOX™ GPL 205,
Datex-Ohmeda stock number 1001-3854-000
Some screws (as noted) require an anti-loosening bond. Use:
• Loctite #24231, screw lock,
Datex-Ohmeda stock number 0220-5016-300
When you replace fittings, position the barb end in the same direction as the
original fitting to make hose connections easier.
8.2 Service tools
Tool
Stock Number
Software Update Kit Using PC, Aestiva
(does not include ventilator software)
1006-8376-000
Cable, PC, Software Update, Aestiva
(included in update kit 1006-8376-000)
1006-3900-000
Software Media, 3.5” disk, Aestiva
Call Technical Support
Power Jumper Cable (provides power to ventilator with outlet box removed from machine)
1006-4014-000
MOPV alignment tool
(included with MOPV elastomers service kit 1503-8017-000)