4a Tests and Calibration 4.X Software
4a-4
05/04 1006-0453-000
4a.2 Service Mode Confirmation menu
The service calibration mode tests and/or calibrates hardware necessary to
prepare a ventilator manufactured for shipment and to service a ventilator in
the field.
There are two ways to enter the service mode:
• If the machine is turned off, push and hold in the adjustment knob while
setting the system switch to On. Hold the adjustment knob pushed in
until the “Service Mode Confirmation” menu appears. Use the
adjustment knob to highlight “Service Mode”, then push the adjustment
knob to confirm the selection.
• If the machine is already in normal operation, set the Bag/Vent switch to
Bag. Press End Case key, press the V
T /
Pinsp, the PEEP, and the menu
switches at the same time to reset the software (powerup). Push and
hold the adjustment knob until the “Service Mode Confirmation” menu
appears.
Figure 4-1 • Service confirmation menu
From this menu you can go to the Service Mode or into Normal Operation.