4a Tests and Calibration 4.X Software
1006-0453-000 05/04
4a-7
4a.3.2 Error Log
There are two special types of alarms:
• Minimum monitoring alarms that stop mechanical ventilation
• Minimum system shutdown alarms that stop mechanical ventilation and
monitoring.
An alarm message that results from these special types of alarms is
considered an error alarm.
The Error Log displays up to 10 of the most current error messages that have
been logged.
Each log entry shows:
•
Bootup Count
- number (the bootup count is incremented each time the
machine is turned on).
•
Time (ms)
- the time in milliseconds since bootup when the latest error
condition occurred.
•
Address
- the place in the software sequence where the last occurrence of
the error took place.
•
Error
- message associated with the particular error condition.
Figure 4-4 • Error Log menu
The error address and software revision are important pieces of information to
note if technical support is required.
The bottom left corner of the screen displays additional information:
• Bootup Count Last Cleared
• Current Bootup Count
Clear Error Log
To clear the error log, select “Clear Error Log”. The system asks you to confirm
that you want to clear the error log.
Remarks
After clearing the error log:
• the Boot Count Last Cleared number will reset to the Current Boot Count
number.
• the menu will show the message “No entries in log!”.
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......