4a Tests and Calibration 4.X Software
4a-8
05/04 1006-0453-000
4a.3.3 System
Configuration
The System Configuration menu includes settings that are tailored to the
specific machine.
Figure 4-5 • Calibrations/System Configuration menu
4a.3.3.1 Altitude
The accuracy of some of the ventilator measurements is altitude sensitive. To
ensure the specified accuracy, the altitude setting should be set to the
specific altitude where each machine is located.
Altitude settings range from -400 to 3600 meters in increments to 100
meters.
Figure 4-6 • Altitude menu item
4a.3.3.2 Drive Gas
Either O2 or Air can be used as the drive gas for the ventilator’s pneumatic
engine. To compensate volume calculations for the specific density of the
drive gas used, the drive gas selection on this menu must match the actual
drive gas.
To change the actual drive gas, refer to section 4.1.9 of the Aestiva Anesthesia
Machine Service Manual.
w
CAUTION
If you change the drive gas, you must also change the drive gas selection
on this service setup screen. If the drive gas selection and the actual
drive gas do not agree, volumes will not be correct.