4a Tests and Calibration 4.X Software
4a-20
05/04 1006-0453-000
4a.3.11 Test Serial Ports
Two serial port tests may be done:
•
Public Port Test (Com 2)
- requires pins 6 and 13 of the serial connector to
be jumpered.
•
Proprietary Port Test (Com 1)
- requires pins 2 and 3 of the serial connector
to be jumpered.
Com 1 and Com 2 ports are located on the back of the ventilator and are
identified with the following symbol:
Each test routine sets up the serial port circuits so transmitted data is echoed
directly back to the receive circuits. The test fails if the data sent out is not
equal to the data received.
If only the Proprietary Port Test fails, replace the CPU.
If both tests fail:
• check the harness connections between the control board and the Serial
Adaptor Board (SAB).
• check the ribbon cable between the SAB and the external connector.
• replace the SAB.