4a Tests and Calibration 4.X Software
4a-24
05/04 1006-0453-000
4a.3.14 Display Discrete
I/O Signals
The
Discrete I/O Signals
menu displays discrete binary signals associated with
machine switch positions.
There are several types of switches in the Aestiva machine:
•
some switches are mechanically operated
•
some switches are pneumatically operated
•
some switches are electronic
Figure 4-25 • Display Discrete I/O Signals menu
Mechanical switches
• ACGO Switch Status — Ventilation or Aux CGO (machines with ACGO)
• CO
2
Bypass — Off or On
• Canister Status — Closed or Open
• Control Panel Cover — Closed or Open
• Bag/Vent Status — Bag or Vent
Pneumatic switches
• Pressure Limit Status — Off or On
• O
2
Flush Status — Off or On
• O
2
Supply Pressure Status — Pressure or No Pressure
Electronic switches
• O
2
Sensor Status — Connected or Disconnected
• Circuit module ID:
Off = tab present to block light
On = no tab to block light
Breathing Circuit
Switch
Circle
Bain/Mapleson D
(no module)
One
Off
On
On
Two
Off
Off
On
Three
On
Off
On