4a Tests and Calibration 4.X Software
4a-28
05/04 1006-0453-000
4a.3.18 Adjust Drive Gas
Regulator
The Adjust Drive Gas Regulator procedure establishes the required flow rate
through the drive gas regulator for proper calibration.
Figure 4-29 • Adjust Drive Gas Regulator menu
Remarks
The drive gas regulator should provide a constant gas input pressure of
172 kPa (25 psi).
You can verify this pressure by attaching a pressure test device to the
regulator pressure port (shown below) and performing the above procedure.
If required, adjust the regulator to 172 ±1.72 kPa (25 ±0.25 psi).
Figure 4-30 • Location of the regulator pressure port
Regulator
pressure
port