4a Tests and Calibration 4.X Software
4a-34
05/04 1006-0453-000
4a.3.24 Service
Calibrations Required
The
Schedule Service Calibrations
menu lists which setting or calibration
must be performed when the “Service Calibration
w
” alarm appears in
normal operation. After the setting or calibration is properly completed, the
text for that setting or calibration is removed from the list.
Figure 4-36 • Service Calibrations Required menu
The normal operation “Service Calibration
w
” alarm message is only removed
when all the required settings or calibrations are completed.
Remarks
The
Set Service Calibration
menu item is used by the
factory
to activate the
Service Calibration alarm and require that all settings and calibrations be
performed when the Aestiva machine is set up for operation at its permanent
location.
There is no need to set the “Service Calibration
w
” alarm in the field.
Select “No” when the following warning appears if you selected the
“Set Service Cal” menu item.
Figure 4-37 • Set Service Calibration menu
If you select “Yes” you will see the screen displayed in Figure 4-38. Follow the
directions on the screen to complete the calibrations.
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......