4b Tests and Calibration 1.X/3.X Software
1006-0453-000 05/04
4b-7
4b.4 Diagnostic Tests
4b.4.1 Test CPU
The software tests the CPU integer instructions as well as the CPU register(s).
If this test fails, the CPU did not perform an integer instruction correctly, or the
CPU register(s) have failed.
CPU test FAIL instructions
Upon FAIL, replace the CPU board (see Section 7.5).
4b.4.2 Test External RAM
The software tests all of the external RAM memory with a walking bit pattern
test. It writes a certain bit pattern to a block of memory and then reads that
block of memory. If the bit pattern that it wrote is not the same bit pattern that
it reads back the test fails.
External RAM test FAIL
instructions
Upon FAIL, replace the CPU board (see Section 7.5).