4b Tests and Calibration 1.X/3.X Software
4b-10
05/04 1006-0453-000
4b.4.7 Test Serial Ports
The software performs an internal test on both of the serial ports. The
software sets up the serial ports so data sent out the serial ports is echoed
directly back to that serial port. The test fails if the data sent out is not equal to
the data received. The status of the test is displayed on the ventilator’s front
screen. The software checks the serial ports to ensure any data sent out is
looped back as a direct echo. If the returned data is not the same as that sent,
the ports test fails.
Serial port test FAIL
instructions
Upon FAIL, replace the CPU board (see Section 7.5).