Page 10
D-Tec Vaporizer
August 2001
O & M Manual Part No. 1107-0622-000
3.
Remove the mains lead retaining plate illustrated on Fig. 11 and fit the Datex-Ohmeda mains lead
firmly into the socket of the vaporizer. Feed the mains lead into the mains lead channel and then clip
the mains lead retaining plate back into the base.
4.
Feed the lead round the back of the vaporizer.
Fig. 11 Fitting The Mains Lead
4.2
Vaporizer Installation
Connecting the vaporizer to the plug-in system
w
Caution: Use only the Dräger plug-in system!
1.
Set the control dial to the
q
position.
On commissioning only:
•
Remove protective plugs from
vaporizer's ports.
2.
Install the ‘O’ rings, if necessary, as
described in the relevant anesthesia machine
operation and maintenance manual.
3.
The locking lever must be set to the “unlocked” position.
•
With both hands, hold the main body of the vaporizer in an
upright position and lower it on to the plug-in system,
ensuring that the vaporizer ports correctly engage the
plug-in system port valves.
•
To lock the vaporizer, turn the locking lever 90
°
in a
clockwise direction.
The vaporizer should lie positioned horizontally on the
plug-in system.
w
Warning: Do not install a D-Tec vaporizer in the right hand
position of a Narkomed 6000 as this will result in an incorrect fit
and may cause leaks in the fresh gas delivery system.
MAINS LEAD
CHANNEL
MAINS LEAD
MAINS LEAD
RETAINING PLATE
AB.54.002
2
2
3
1
AB.54.010
AB.54.009