D-Tec Vaporizer
Page 11
O & M Manual Part No. 1107-0622-000
August 2001
Fitting the equi-potential lead
e.g. for intracardiac
•
Fit one end of the lead to the equi-potential stud on
the D-Tec.
•
Route and connect the other end of the lead to a
central ground stud, as described in the relevant
anesthesia machine Operation and Maintenance
Manual.
DESFLURANE
NO OUTPUT
LOW AGENT
ALARM
BATTERY LOW
WARM-UP
OPERATIONAL
5.0
Operating Instructions
5.1
Alarm And Display Test
w
Warning: Only use vaporizers which are in a serviceable condition.
w
Caution: Allow the vaporizer to attain its specified ambient operating temperature before connecting it
to a mains power supply.
1.
Connect the mains lead from the vaporizer to an approved hospital grade outlet socket.
w
Warning: Do not use the vaporizer if during the alarm and display test any one of the five lights and all
the LCD agent level indicator bars do not flash or the auditory alarm does not operate for a period of
approximately two seconds.
2.
Verify that the alarm and display test is automatically operated for a period of approximately two
seconds as follows:
a)
Each light and all the LCD agent level
indicator bars on the front display
panel flash four times.
b)
The auditory alarm is activated four
times.
3.
The alarm and indicators can be tested at any time
by pressing the
for at least four seconds to
activate the alarm and display test. Activating the
test does not affect the operation of the vaporizer.
4.
The alarm and display test is designed to test the
lights, the LCD bars and the auditory alarm. If any
of the lights, the LCD bars and/or the auditory
alarm fail to operate, do not use the vaporizer.
Fig. 12 Alarm And Display Test
AA.43.048
AB.54.005