7 Troubleshooting
1505-1018-000 08/07
7-21
Cannot calculate
FRC
FRC invalid
Low
An FRC measurement was started for
FRC or PEEP Inview, but unable to be
completed within the last 5 seconds.
• Check end tital CO2 reading.
• Ensure airway module is available and active.
• Stop any active procedures.
• Start a new FRC measurement.
• Ventilation changes will stop an FRC measurement.
• With gas module plugged in, check power supply status LEDs on MIB from bottom of sheetmetal enclosure.
If any LEDs are lit, then unplug module. If LEDs remain lit, replace MIB.
• Replace gas module.
Check D-fend
CO2 or O2 wave
instruction:
Check D-fend.
Wait 30 sec and
press
Normal
Screen
to
continue.
MGAS D-fend Not
Connected >40 Sec
MGAS D-fend Not
Connected >60 Sec
Low
Medium
MGAS communicates that the D-fend
module is not installed.
The pump turns Off.
If
Normal Screen
is
selected the DU commands
the MGAS to start sampling
again.
• Check that the sample tube is connected.
• Check that the water trap is connected.
• Check that the gas outlet is not blocked.
• Replace Module.
Check sample
gas out
CO2 or O2
waveform
instruction:
Check sample gas
out. Wait 30 sec
and press
Normal Screen
to continue.
MGAS Check Sample
Gas Out > 40 Sec
MGAS Check Sample
Gas Out > 60 Sec
Low
Medium
MGAS communicates continuous
occlusion for 40 seconds.
MGAS communicates continuous
occlusion for 60 seconds.
The pump turns Off.
If
Normal Screen
is
selected, the DU commands
the MGAS to start sampling
again.
• Check the gas outlet.
• Replace module.
Clean neo flow
sensor
Neo Flow Sensor
Contaminated
Medium Neo flow sensor reporting flow > 32 l/
min for 4 continuous seconds as
measured by the VCB or VMB.
• Calibrate neo flow sensor.
• Clean neonatal flow sensor
• Replace neo flow sensor.
• Replace neo flow sensor cable.
• Replace VMB.
• Repace motherboard.
• Replace VCB.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...