Volume delivery
error
Total Flow Sensor
Communications
Failure
High
Out of range total flow sensor data
-or-
Loss of communication.
Return to closed loop volume
delivery control when cleared.
• Check Error Log for Total Flow Sensor Supply Voltage.
-If yes:
- Disconnect harness at flow sensor. If voltage error stops, replace flow sensor.
-Disconnect vent engine harness at VCB. If voltage error stops, replace harness.
• If no:
-Check connection to Total Flow Sensor.
-Interchange harness connection between air and total flow sensor.
-If this alarm stops, replace total flow sensor. If this alarm persists, check/replace harness. If harness is OK,
replace VCB.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting