7 Troubleshooting
1505-1018-000 08/07
7-39
7.7 Troubleshooting Service App messages
The VCB Service App screen includes a
Calibrations and Tests
section (Section 8.6.5).
This section includes a field “
Calibration Running
” that displays messages while a test
is in progress. Some messages indicate that the test encountered a problem. Use the
following table to help troubleshoot messages that result in a failed test or calibration.
Message in
“
Calibration running”
field
Reason
Troubleshooting
120 Lpm flow test
Prior to calibration flow valve, make sure system
can flow 120 Lpm for at least 1 second.
None
125 Cal Point Too High:
Value
Exhalation valve calibration point at 1.25
cmH2O was above 1.30 cmH2O.
Value = pressure (cmH2O x 100) recorded at
the 1.25 cm H2O cal point.
Verify exhalation housing and assembly are
seated properly.
300 point too high
Flow valve calibration point targeting 300
mLpm was measured above 290 mLpm and
could not be extrapolated down.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine system is not
leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Aborted
Calibration or test in progress has been
aborted.
None
Air Flow Sensor Chk
Air Flow Sensor Chk is in progress.
None
Air Flow Sensor Chk
Complete
Air Flow Sensor Check passed.
None
Air Flow Sensor Chk Fail
Air Flow Sensor Check failed.
See previous message in status log for cause of
failure.
Air Flow Sensor Fail: No Air
Air Flow Sensor step of Checkout has failed
because no low/no air supply pressure was
detected.
Verify proper supply pressure.
Air Flush Time: Value sec
Value = time (sec) needed to flush system with
Air until stability of 0.1% O
2
is reached.
None
Air Valve Cal
Air Valve Cal in progress.
None
Air Valve Cal Complete
Air Valve Cal passed.
None
Air Valve Cal Fail
Air Valve Cal failed.
See previous message in status log for cause of
failure.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...