7 Troubleshooting
1505-1018-000 08/07
7-53
Span
Pressure transducer span calibration is in
progress.
None
Start Normalization...
Expiratory flow sensor normalization is in
progress.
None
T1 =
Intermediate value in low pressure leak check
compliance calculation.
None; for information only.
T2 =
Intermediate value in low pressure leak check
compliance calculation.
None; for information only.
Target flow not reached @
300
Flow remains greater than the target flow with
the minimum flow command during Checkout.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Total:
Value1
O2:
Value2
During O
2
flow sensor check, O
2
flow sensor is
more than 15% different than total flow sensor
at a nominal flow of 30 L/min.
Value1
= measured total flow.
Value2
= measured O
2
flow.
1. Check regulator for oscillation. Replace if
necessary.
2. If only O
2
flow sensor check fails, replace O
2
flow sensor. If both Air and O
2
flow sensor
checks fail, replace total flow sensor.
unbounded Qin = Value
Intermediate value in low pressure leak check
compliance calculation.
None; for information only.
Upstream Flow Diff: Value
Difference between flow measured by total flow
sensor and inspiratory wire of neonatal flow
sensor during Checkout (mLpm).
None; for information only.
Upstream Flow Failed
(Upstream)
The inspiratory flow measured by the neonatal
flow sensor is more than 15% different from the
flow measured by the total flow sensor.
Refer to Section 7.1.11
Valve Step Sweep
Valve step sweep in progress.
None
Valve Step Sweep Aborted
Valve step sweep aborted because maximum
flow was reached.
None; normal operation.
Valve Step Sweep Done
Valve step sweep completed.
None
Valve Sweep
Valve sweep in progress.
None
Valve Sweep Done
Valve sweep completed.
None
VMB Exp Flow Sensor Cal
Fail
Expiratory flow sensor calibration has failed.
Verify expiratory flow sensor and exhalation
housing are properly installed.
Message in
“
Calibration running”
field
Reason
Troubleshooting
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...