8 Software Download and Diagnostics
1505-1018-000 08/07
8-21
8.6.9 Logging Calibration
Results
The ventilator must be in the Service Mode to log the calibration results.
• Power on the system and wait for the Patient Setup menu to display.
Note: For systems with 4.2X software, wait for the Select Patient menu to
display, then select
Patient Setup menu
.
1. Access the Service Calibration menu.
2. Connect a PC to the system and start the PC Application software (Revision
1.58 or higher).
3. Access the
Log Calibration result to file
area of the PC Service Application
and select
Find Log File
.
4. Select the desired folder from the PC hard drive to store Log Files.
• Rename the file and select
OPEN
.
Suggested file naming convention: Serial date+"cal.log"
• If a Log File does not exist a message window displays. "Select
Yes
to
create a Log File."
5. Verify the file address location displayed in the
Log Calibration results to
file area
and select
Start Logging
.
• The Status Log window displays and indicates that the log has been
initialized.
6. From the EC Ventilator, access the
Calibrations menu
from the
Install/
Service
menu.
• Select each calibration and follow the screen instructions to complete
the calibrations.
• From the PC, select
Save Log
from the Status Log window after
calibrations are complete. Select
Ignore
if "Debug Assertion Failed"
message appears.
As each system calibration occurs the Status Log will display Pass or Fail. If a
calibration fails the log will record and display information about the exact
parameter that failed. This information is essential in determining the cause
for calibration failure(s).
Always include the Log file when iTrak quality
reports are submitted.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...