1 Introduction
1505-1018-000 08/07
1-7
1.6 Display controls and indicators
Figure 1-4 • Controls and indicators
2
4
3
5
3
3
7
1
6
8
AB.98.139
1
Alarm LEDs
The red and yellow LEDs indicate the priority of active alarms.
2
Silence Alarm key
Push to silence any active, silenceable high and medium priority alarms or
to suspend any non-active high or medium priority alarms. Alarm audio is
silenced or suspended for 120 seconds for Adult and Pediatric patient
types, and for 30 seconds for Neonatal patient types. Push to clear resolved
alarms.
3
Menu keys
Push to show corresponding menu.
4
ComWheel
Push to select a menu item or confirm a setting. Turn clockwise or
counterclockwise to scroll menu items or change settings.
5
Normal Screen key
Push to remove all menus from the screen.
6
AC mains indicator
The green LED lights continuously when the EC is connected to an AC mains
source. The internal batteries are charging when the LED is lit.
7
Quick keys
Push to change corresponding ventilator setting. Turn the ComWheel to
make a change. Push the Quick key or ComWheel to activate the change.
8
!
O2 key
Push to deliver 100% O
2
(Adult or Ped) or 25% above current setting (Neo)
for 2 minutes.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...