Engström Carestation
4-18
08/07 1505-1018-000
4.4.8 Resetting
feature option key
codes
1. Power up the Engström Carestation and select
Service
Menu
-
Options List
.
2. Record the available options that are "On".
3. Power down the ventilator.
4. Perform the software download per TRM Section 8.4,
"Download New"
.
5. After software download is complete, power up the ventilator and select
Service
Menu-Options List
.
6. Verify the options "On" match the options recorded in step 2. (IMPORTANT)
• If the "On" options match the options recorded in step 2, proceed to step 11.
• If the options "On" do not match the options recorded in step 2, go to the Datex
Ohmeda "Key Code Generator" Web site: www.docodes.com, and follow the
instructions to request a new "Key Code".
• After you receive the new
Key Code
, select
Service Menu
-
Options Key
. Enter the
new key code, select
Save New Key
and proceed to step 7.
7. Select
Service Menu- Options key
. Record the new three digit Control Board ID and
new seven digit Key Code on the new Key/BID label. (1504-3521-000)
8. Remove the existing Key/BID label from the back of the machine.
9. Clean the area with Isopropyl alcohol.
10.Affix the new Key/BID label to the machine. (1504-3521-000)
11.Perform
Super User Calibrations
per TRM Section 5.1.1.
12.Perform Checkout procedure (
Section 3)
.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...