Engström Carestation
5-2
08/07 1505-1018-000
5.1 Calibration (super-user)
The EC ventilator includes integrated software that allows a qualified user to
periodically calibrate the O
2
flow valve, the Air flow valves, and the Exhalation
valve.
At a minimum, these calibrations should be completed every 6 months,
whenever performance is questioned, or when associated components are
serviced or replaced. These calibrations can be performed more frequently,
as needed, for optimal performance.
The menu structure for accessing the calibration software is described in
Section 4, “
Installation and Service Menus
” .
Note
The Valve calibrations (Section 5.1.1) can be performed without performing
the Service level test and calibrations (Section 5.2). However, if the Service
level test and calibrations are performed, the Valve calibrations should be
performed again after the Service level test are completed (passed).
Note
To file a quality report or request technical assistance regarding a calibration
failure, create and include a Calibration log while using the EC Service
application .
Run the Calibration while the EC Service Application is running and select
Log
Calibration Results .