5.2 Service level tests and calibration
The service level tests require the Windows based service application as
described in Section 8.5, “
EC Service Application (PC based)
” .
Note
After completing the Service level test and calibrations, perform the Valve
calibrations in Section 5.1.1.
5.2.1 Service application
setup
1. Connect a Windows based PC to the serial port (Port 3) of the EC using a
USB to RS-422 converter (Refer to section 10.1.1).
2. Start the Service Application on the PC.
• Open the
VCB
and
VMB
windows.
3. To establish communication between the PC Service Application and the
EC, the EC must be powered On.
a. Read only access is available in normal operation mode.
b. Controlled access is available in Super User or Service modes.
5.2.2 Test setup
1. Connect a manual shut-off valve to each gas supply inlet.
• Remove the inlet fittings from the EC and ensure o-ring remains on the
adapter).
• Connect the shut-off valve to the respective EC pipeline adapter.
• Connect the removed inlet fitting to the respective shut-off valve.
2. Close both shut-off valves.
3. Connect pipeline supplies to each shut-off valve.