Datex-Ohmeda S/5 monitors
Document No. 800 1009-1
20
CPU Board
Lamp Unit
OM Board
PVX
valves
MUX
Valves
A / D
M
o
d
u
le
B
u
s
TPX
Lamp
Gas press
Diff press
MUX &
buffer
Valves
CPU
80C196NT
Fan
pneumatics
unit
reset
data
Lamp
Current
Sensing
A / D
Temp
sensor
Figure 13
Control logic
CPU Board
TPX Board
PVX Board
OM Board
M
odu
le
B
u
s
RS-485
driver
CPU
80C196NT
EEPROM
Factory calibration
data for PVX
Factory calibration
data for OM and TPX
EEPROM
Correction data for
amplification and
offset of all amplification
channels
EEPROM
EEPROM
Figure 14
Calibration data stored in EEPROM
2.2.7
OM board
The Oxygen board contains the specific electronics for the oxygen sensor. Sample flow
measurement and sampling system pressure sensors are on this board. It also contains EEPROM’s
that store calibration data of both TPX and OM-sensors. The spirometry keyboard connection is on
this board.
2.2.8
PVX board
The Spirometry board is connected to the oxygen board. It contains pressure sensors for airway
pressure and flow measurement differential pressure and preamplifiers for those. Calibration data
of spirometry is stored on it’s own EEPROM.