7 Specifications

1175-0013-000

7-5

Sevotec 5%

Figure 7-5 • Effect of Flowrate at 21 ± 2° C with oxygen flowing

Figure 7-6 • Effect of temperature at 5 liters/minute with oxygen flow

AB800031

1

2

3

4

5

0.2

1

5

10

5

4

3

2

1

0.6

0.2

15

0

% SEVOFLURANE

DIAL SETTING

Flowrate (liters/minute oxygen)

AB80032

0

1

2

3

4

5

15

20

25

30

5

4

3

2

1

0.6

0.2

35

% SEVOFLURANE

DIAL SETTING

Temperature °C

Summary of Contents for Tec 7

Page 1: ...Tec 7 Vaporizer User s Reference Manual ...

Page 2: ...n accordance with written instructions provided by Datex Ohmeda and by Datex Ohmeda trained personnel The Product must not be altered without the prior written approval of Datex Ohmeda The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone other than Datex Ohmeda wCAUTION U S ...

Page 3: ...ounting procedure 3 1 Mounting the vaporizer 3 2 Checking the vaporizer for correct mounting 3 4 Removing the vaporizer from a manifold 3 4 4 Operating Instructions Setting the dial 4 1 Filling and draining the vaporizer 4 3 Filling procedure with funnel filler 4 5 Draining procedure with funnel filler 4 6 Filling procedure with Easy Fil 4 7 Draining procedure with Easy Fil 4 8 Filling procedure w...

Page 4: ... 5 6 Principle of Operation Interlock mechanism 6 1 Delivery of gas agent vapor 6 1 Overview 6 1 Bypass circuit 6 2 Vaporizing chamber circuit 6 2 7 Specifications Calibration 7 1 Performance 7 2 Weight and dimensions 7 2 Flow characteristics 7 3 Effects of variables 7 8 Anesthetic agent consumption 7 8 Barometric pressure 7 8 Ambient temperature 7 9 Back pressure 7 9 Carrier gas composition 7 10 ...

Page 5: ...n the order of a licensed medical practitioner Outside the U S check local laws for any restrictions that may apply Do not attempt to use a vaporizer that has been dropped A dropped vaporizer MUST be sent to the nearest Datex Ohmeda Field Operations Unit for servicing Do not use malfunctioning equipment Make all necessary repairs or have the equipment serviced by an authorized Datex Ohmeda service...

Page 6: ...ion including this manual and accompanying labels immediately available to all users Symbols Warnings and Cautions tell you about conditions that can occur if you do not follow all instructions in this manual w w w w WARNING Warnings tell about a condition that can cause injury to the operator or the patient w w w w CAUTION Cautions tell about a condition that can cause damage to the equipment Rea...

Page 7: ... compensated so that its output remains relatively constant despite cooling due to evaporation variations in inlet flow and fluctuating pressures as described in Section 7 Effects of Variables The vaporizer is designed to be used on Selectatec Series Mounted Manifolds The vaporizer can be installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec Series ...

Page 8: ...ance of the anesthesia machine and vaporizer can be degraded if the machine and vaporizer are mis matched Only operate the vaporizer with dry medical gases Ifavaporizercontainingagentinthesumphasbeeninverted connectit to a gas scavenging system set the dial to 5 and purge the vaporizer with the carrier gas at 5 liters minute for 5 minutes w w w w CAUTION The vaporizer is intended to be operated in...

Page 9: ...wo or more vaporizers are fitted on a Selectatec Series Mounted Manifold The gas flow enters only the vaporizer that is turned ON Any unwanted anesthetic trace vapor is minimized after a vaporizer is turned to w WARNING Earlier versions of the Selectatec Series Mounted Manifold that provide mountingpositionsforthreevaporizersrequirethatifonlytwovaporizers are fitted then the center position must b...

Page 10: ...NING Do not affix any additional labels or markings to the back panel They may adversely affect the operation of the vaporizer identification unit 1 Vaporizer identification label Figure 2 2 Vaporizer identification label HALOTHANE AB80011 1 ...

Page 11: ...to the Selectatec Series manifold ensure that each manifold port valve O ring is intact and that there is no foreign matter around the mating surfaces A damaged O ring and or foreign matter around the mating surfaces can cause leaks Earlier versions of the Selectatec Series Mounted Manifold that provide mountingpositionsforthreevaporizersrequirethatifonlytwovaporizers are fitted then the center po...

Page 12: ...ve any plugs fitted to the vaporizer interlock block ports Verify that each manifold port valve O ring is intact If necessary remove the existing O rings and fit one new O ring to each port valve as described in the relevant anesthesia system User s Reference Manual Replacement O rings are supplied with each vaporizer 1 Vaporizer Interlock Block Port ensure plugs removed 2 Replace Manifold Port Va...

Page 13: ... with both hands Lower the vaporizer onto the manifold ensuring that the vaporizer interlock block ports engage correctly with the manifold port valves Figure 3 4 Installing the vaporizer Step 5 Lock the vaporizer onto the manifold Push the locking lever all the way down Turn it clockwise to the locked position to lock the vaporizer onto the manifold 1 Locking lever Figure 3 5 Locking the vaporize...

Page 14: ...o be level and the locking lever is in the locked position attempt to lift the vaporizer straight up from the manifold If the vaporizer can be lifted off the manifold it is not correctly mounted Remount the vaporizer see Vaporizer mounting procedure Verify that the interlock rods are in alignment by making sure that only one vaporizer at a time can be turned ON Check the anesthesia system for leak...

Page 15: ...ease the locking lever and check that the locking lever springs up to the unlocked position to release the vaporizer from the manifold 1 Locking lever 2 Dial Figure 3 7 Unlocking the locking lever Step 3 Carefully lift the vaporizer up from the manifold Figure 3 8 Lifting the vaporizer AB80003 Unlocked 1 2 1 2 Locked AB80019 ...

Page 16: ...Tec 7 Vaporizer 3 6 1175 0013 000 ...

Page 17: ...l if the vaporizer is not properly locked onto the manifold Important European Standard EN 740 Anesthetic Workstations and Their Modules requires that an appropriate gas monitor is used to monitor the concentration of anesthetic agent vapor in the inspiratory gas when the vaporizer is in operation in order to provide protection against hazardous output in the event of a device malfunction Datex Oh...

Page 18: ... that it is not possible to turn on the vaporizer if an adjacent Tec series except Tec 3 is turned on Step 2 The vaporizer should not be used between and the first graduation mark To avoid inadvertent delivery of small concentrations turnthecontroldialto when the vaporizer is not in use Figure 4 1 Releasing the dial Åben AB80020 ...

Page 19: ... flammable anesthetic agents in the IEC 60601 2 13 Standard Particular Requirements for the Safety of Anesthesia Machines are suitable for use in the presence of this vaporizer Asthisvaporizerisnotsuitableforusewithflammableanestheticagents suchasEtherorCyclopropane theuseofantistaticbreathingtubesand face masks is not necessary The use of antistatic or electrically conductive breathing tubes when...

Page 20: ...used when the anesthetic agent does not contain additives or stabilizing agents but the procedure must be performed at least once every year The following steps should be taken for Halothane vaporizers Drain the vaporizer every two weeks If Halothane is used infrequently the vaporizer should be drained after use The decomposition of halothane causes the release of halides which may corrode metal c...

Page 21: ... it with the tool end of the filler cap Figure 4 3 Filling a vaporizer that incorporates a funnel filler Figure 4 4 Replacing the filler cap Step 2 Verify that the anesthetic agent is the same as that specified on the vaporizer front label Observe the agent level through the sight glass indicator on the side of the filler body Pour the agent slowly into the filling port as illustrated on Fig 4 3 u...

Page 22: ...ug below the filling port on the filler body as illustrated in Fig 4 5 Figure 4 5 Draining a vaporizer with a funnel filler Figure 4 6 Replacing the filler cap Step 2 Position a properly marked container under the drain spout Step 3 Unscrew but do not remove the drain plug to allow the vaporizer contents to pour from the drain spout into the container Step 4 After draining is complete tighten the ...

Page 23: ...8 Replacing the filler cap Step 2 Remove the filler cap Align the bottle adapterkeyswiththeindexslotsinthe filler block as illustrated in Fig 4 7 Insert the bottle nozzle into the filler block Step 3 Press the agent bottle fully into the vaporizer filler block Allow the liquid to flow into the vaporizer until the maximum level mark is reached To help prevent overfill ensure the agent level is at o...

Page 24: ...Remove the cap from the vaporizer filler Figure 4 9 Unscrewing the drain plug Figure 4 10 Replacing the filler cap Step 2 Place the empty container opening under the drain nozzle as shown in Fig 4 9 Step 3 Unscrew the drain plug with the tool attached to the filler cap Drain the vaporizer until empty Step 4 Afterdraining tightenthedrainplugto help minimize leaks Step 5 Replace and tighten the fill...

Page 25: ...block Rotate the bottle to align the bottle filler nozzle keys with the index slots in the filler block as illustrated on Fig 4 11 Step 3 Press the agent bottle fully into the vaporizer filler block Allow the liquid to flow into the vaporizer until the maximum level mark is reached Pay particular attention to the level in the sight glass and the air return bubbles flowing into the bottle Step 4 Re...

Page 26: ...ap from the vaporizer filler Figure 4 13 Unscrewing the drain plug Figure 4 14 Replacing the filler cap Step 2 Place the empty container under the drain nozzle as shown in Fig 4 13 Screw bottle onto drain spout Step 3 Unscrew the drain plug with the tool attached to the filler cap Drain the vaporizer until empty Step 4 Afterdraining tightenthedrainplugto help minimize leaks Step 5 Replace and tigh...

Page 27: ...or Halothane vaporizers check the output of anesthetic agent periodically with an agent monitor See note below Note The decomposition of Halothane causes the release of halides which may corrode metal components particularly in the presence of moisture Also a preservative added to Halothane by its manufacturers to impede decomposition can leave a residue which may cause vaporizer components to sti...

Page 28: ...Remove the vaporizer from service immediately and label the vaporizer stating that it is contaminated Discard all liquid 2 Return the vaporizer to a Datex Ohmeda Authorized Service Center stating that the vaporizer is contaminated and if possible the type of contaminant in the vaporizer Output concentration check Connect the Tec 7 to an Anesthesia Machine 1 Set the oxygen output of the anesthesia ...

Page 29: ... of performance have developed 1 To predict the concentration that the vaporizer can be expected to deliver the detailed nominal performance data and the preceding comments must be taken into account 2 The method of test used must not be such that it bears little relation to normal conditions of clinical use 3 Any sampling techniques used must be such as to ensure the following a The sample is ful...

Page 30: ...the calibration and reliability of analyzers must be realistically considered and account must be taken of errors of use The following method of checking can be used where special equipment is not available or where a secondary check of analyzers is desired The characteristics of the vaporizer are such that if the vaporizer is satisfactory at one dial setting it should be satisfactory at all other...

Page 31: ...st be taken to handle the exhaust gases and spillage Note This method is designed to be a quick and easy check of vaporizer operation and therefore it is somewhat imprecise However it is unusual for the measured liquid consumption to vary by more than 25 of the values listed above Service Policy Repairs and service procedures must be performed at a Datex Ohmeda Authorized ServiceCenter Contactyour...

Page 32: ...Tec 7 Vaporizer 5 6 1175 0013 000 ...

Page 33: ...equence which allows the dial release to move out to lock the dial in the position closes the manifold port valves and vents the vaporizer gas connecting ports and allows an adjacent vaporizer to be turned ON Turning the locking lever to the unlocked position releases the vaporizer allowing it to be removed from the manifold Delivery of gas agent vapor Overview The output concentration of the Tec ...

Page 34: ...or channel 9 Rotary valve Figure 6 1 Bypass circuit Vaporizing chamber circuit The fresh gas flow through the vaporizing chamber as shown on Figure 6 2 flows from the flowmeter across the sump cover where it is diverted through the central cavity of the rotary valve and back through the Intermittent Positive Pressure Ventilation IPPV compensating assembly Gas now flows from the IPPV assembly down ...

Page 35: ...3 Combined fresh gas and enriched gas out 4 Fresh gas bypass 5 Fresh gas out 6 Thermostat 7 Vaporizing chamber 8 Wick assembly 9 IPPV compensating assembly 10 Sump cover 11 Vapor control channel 12 Shown in ON position Figure 6 2 Vaporizer schematic 11 1 2 3 4 5 6 7 8 9 AB80005 10 12 ...

Page 36: ...Tec 7 Vaporizer 6 4 1175 0013 000 ...

Page 37: ...tion certificate that is included with your Tec 7 Vaporizer Vaporizers are calibrated at 21 C using an oxygen carrier gas at a flow of 5 liters minute and they are temperature flow and pressure compensated within the specified operating range w WARNING The Tec 7 Vaporizer can only be calibrated at a Datex Ohmeda Authorized Service Center ...

Page 38: ...nal Retained by wick system 75 ml nominal To fill from minimum to maximum mark 5 Vaporizer 170 ml nominal 8 Vaporizer Sevotec 137 ml nominal Flow resistance at 5 liters minute of O2 at 21 2 C 10 15 cm H2O with Vaporizer setting ON 0 Operating temperature range 18 C to 35 C 64 F to 95 F Storagetemperature range 40 C to 65 C 40 F to 149 F Note Protect the vaporizer packaging from condensation Weight...

Page 39: ... at 21 2 C with oxygen flowing Figure 7 2 Effect of temperature at 5 liters minute with oxygen flow 0 1 2 3 4 5 0 2 1 5 10 15 5 4 3 2 1 0 6 0 2 AB80027 ISOFLURANE DIAL SETTING Flowrate liters minute oxygen 0 1 2 3 4 5 15 20 25 30 5 4 3 2 1 0 6 0 2 35 AB80028 ISOFLURANE DIAL SETTING Temperature C ...

Page 40: ... with oxygen flowing Figure 7 4 Effect of temperature at 5 liters minute with oxygen flow 0 1 2 3 4 5 0 2 1 5 10 15 5 4 3 2 1 0 6 0 2 AB80029 HALOTHANE DIAL SETTING Flowrate liters minute oxygen 0 1 2 3 4 5 15 20 25 30 5 4 3 2 1 0 6 0 2 35 AB80030 HALOTHANE DIAL SETTING Temperature C ...

Page 41: ...ith oxygen flowing Figure 7 6 Effect of temperature at 5 liters minute with oxygen flow AB800031 1 2 3 4 5 0 2 1 5 10 5 4 3 2 1 0 6 0 2 15 0 SEVOFLURANE DIAL SETTING Flowrate liters minute oxygen AB80032 0 1 2 3 4 5 15 20 25 30 5 4 3 2 1 0 6 0 2 35 SEVOFLURANE DIAL SETTING Temperature C ...

Page 42: ... C with oxygen flowing Figure 7 8 Effect of temperature at 5 liters minute with oxygen flow 1 2 3 4 5 0 2 1 5 10 5 4 3 2 1 0 6 0 2 15 0 AB80035 ENFLURANE DIAL SETTING Flowrate liters minute oxygen 0 1 2 3 4 5 15 20 25 30 5 4 3 2 1 0 6 0 2 35 AB80036 ENFLURANE DIAL SETTING Temperature C ...

Page 43: ...lowing Figure 7 10 Effect of temperature at 5 liters minute with oxygen flow 1 0 2 1 5 10 5 4 3 2 1 0 6 0 2 15 0 2 3 4 5 6 7 8 6 7 8 AB80033 SEVOFLURANE DIAL SETTING Flowrate liters minute oxygen 1 15 20 25 30 5 4 3 2 1 0 6 0 2 35 0 2 3 4 5 6 7 8 6 7 8 AB80034 SEVOFLURANE DIAL SETTING Temperature C ...

Page 44: ...king figure 1 milliliter of liquid anesthetic is required to provide 200 milliliters of vapor The approximate hourly consumption of anesthetic agents can be expressed as 3 3 x x F where represents the setting of the vaporizer output percentage and F represents the input flowrate in liters minute Example If a vaporizer is set to deliver 2 at 6 liters minute total gas input flowrate then the approxi...

Page 45: ... equation given in the section Barometic Pressure therefore applies with the term P now being the absolute pressure at the outlet that is barometric pressure plus back pressure Steady back pressure reduces the v v Currently it is unlikely that the steady back pressure imposed by commonly used downstream components other than some ventilators exceeds 30 mmHg at commonly used flowrates Back pressure...

Page 46: ...This is a normal machine characteristic and is caused by residual vapor left in the machine from previous use When the vaporizer is turned from the setting after a period out of use a brief high concentration may occur that rapidly stabilizes to the set concentration within approximately 10 seconds at 5 liters minute These phenomena are normal characteristics of vaporizers In use the volume of vap...

Page 47: ...to expendable parts The foregoing warranties shall not apply if the Product has been repaired other than by Datex Ohmeda or in accordance with written instructions provided by Datex Ohmeda or altered by anyone other than Datex Ohmeda or if the Product has been subject to abuse misuse negligence or accident Datex Ohmeda s sole and exclusive obligation and Buyer s sole and exclusive remedy under the...

Page 48: ...786 7420 M M M Ma a a all l la a a ay y y ys s s sii i ia a a a Datex Ohmeda Pte Ltd Level 2 Bangunan O Connor 13 Jalan 223 46100 Petaling Jaya Selangor West Malaysia Tel 60 3 754 7872 Fax 60 3 757 6948 S S S Sii i in n n ng g g ga a a ap p p po o o or r r re e e e Datex Ohmeda Pte Ltd 152 Beach Road 12 05 07 Gateway East Singapore 189721 Tel 65 391 8618 Fax 65 291 6618 T T T Th h h ha a a aii i i...