vPad-A1 OPERATORS MANUAL
Cardiac Output:
• Baseline Temperatures: 36, 37 and 38°C.
• Injectate Thermistor Types: - H-P/Philips
- American Edwards/Spacelabs
- Adjustable (default Edwards/Spacelabs at 20°C)
• Injectate Temperatures: 0, 2, 20 or 24°C. Note: 24°C Spacelabs is adjustable
• Cardiac Outputs: 3, 4, 5, 6, and 7 liters/minute
• Simulation Types: - Slow Injection
- Faulty Injection
- L/R Shunt
- Cal Pulse: 1°C for 1 second
Fetal ECG/Intrauterine Pressure:
• Fetal ECG Rates: 60 to 240 BPM; 1 BPM increments
• Fetal ECG Amplitudes: See NSR specification
12 preset rates, user definable
• Fetal ECG Rate Variation: Uniform, Early, and Late Deceleration; Uniform
Acceleration
• Pressure Waveform: Bell shaped, 0-50 or 0-90 mmHg
• Contraction Frequency: Every 2, 3, 5 minutes, or manually triggered
• Transducer Sensitivity: 5 µV/V/mmHg or 40 µV/V/mmHg
Auto Presets:
Unlimited number of user-defined simulation setups can be programmed into the vPad-A1.
Specifications/Chapter 1
#
Page 8