vPad-A1 OPERATORS MANUAL
4.8.2.3 Calibration: SpO2 R-Curves
In a typical pulse oximeter, the
SpO2 R-Curve
is the means by which that
particular device translates its measurement of light transmission through
oxygenated blood to an oxygen saturation in percent. Individual
R-Curves
are derived through clinical studies and, while similar, may be different for
each make and model encountered. vPad-A1 has
R-Curves
pre-
programmed for a number of common monitor makes and models;
however, users may need to add a new device to the make/model menus.
The
button in the
Settings
panel launches
a calibration tool for
SpO2 R-Curves
. The
Manufacturer
,
Model
, and the
(optional)
Sensor
field in the calibration tool uniquely identify each
individual
R-Curve
.
Note
: The
Sensor
identifier accounts for small variations in output that can exist
between an OEM sensor (e.g. Nellcor's DS100) and a third-party
replacement sensor (e.g. EnviteC).
Users can enter new information into the three text input boxes as
appropriate for the new device. Alternatively, press the
Folder Search
icon to choose an existing
R-Curve
as template. A new window will show a
folder list of known
Manufacturers
, as in
Figure 97
.
Touching any
Manufacturer
folder will populate the
Models
dropdown menu
with existing devices in the same category. Once the desired
Model
is chosen
from the list, users can confirm the selection by pressing
.
If a template
R-Curve
was successfully selected, the
Manufacturer
,
Model
, and
(optional)
Sensor
fields will now be populated with text from the template,
along with all additional data fields discussed below. Users should edit the
auto-filled text to create a new device
R-Curve
; otherwise, the template may
be over-written with the newly-entered data.
Having identified the oximeter device, the following steps will create for it a suitable
R-Curve
.
Optimize Signal Strength
•
When the pulse oximeter sensor is placed on the vPad-O2 finger, the
Signal Strength
indicator will display the output of the oximeter's
sensor in real time.
Figure 96
Figure 97
Figure 98
Manual Operation/Chapter 4
#
Page 67
Summary of Contents for vPad-A1
Page 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Page 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Page 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Page 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Page 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Page 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Page 108: ......