Safety Notes
Installation / Update
New installations, updates as well as major interventions on the configuration
of the ergo_bike medical8 may only be carried out by authorised personnel.
Potential Equalisation
Connecting of electronic devices through V 24 / RS232 / USB / LAN, for instance,
or other interfaces of a computer, or of another independent measuring device
may only be carried out by authorised personnel and in rooms intended for
medical use according to DIN VDE (German Electrotechnology Federation)
0100 / Part 710.
Parallel connection over potential equalisation makes sure that in the case of
an eventual insulation breakdown none of the networked devices will develop
on its housing a voltage in excess of 10mV, and thus harm the patient. For this
purpose all the medical devices of daum electronic gmbh are equipped with a
connector for potential equalisation; this connector must be connected by means
of a potential equalisation cable to all the other devices in this system and to
the potential equalisation bus of the room.
Other Notes
All the diagrams and evaluations are to be considered as a suggestion from the
device. When providing a diagnosis the results must always be controlled and
evaluated by a qualified physician.
The electronic installations of the rooms where the device will be used must
absolutely comply with the requirements of directive DIN VDE (German
Electrotechnology Federation) 0100 / Part 710.
The device is not suited for operation in rooms and / or areas with explosion
risks.
All the interventions in the device, modifications to system components, extensions
to the system, as well as needed internal cleaning and repairs may only be
carried out by personnel authorised by daum electronic gmbh.
When using the device you must absolutely follow (respect) the instructions in
the user manual for the load-applying unit, particularly with respect to the safety
of the patient.
Cellular telephones may not be used in the direct vicinity of the device, otherwise
they may interfere with its functionality.
Wireless telephones may be used.
Electromagnetic interference are primarily characterised by the oscillation of the
displayed load value. If the displayed value varies uncontrollably frequently,
even when the speed is higher than 30 /min, this can be an effect of electromagnetic
compatibility (EMC).
Only cables certified by daum electronic gmbh may be used to connect the
device with other devices.