Safety Notes

Installation / Update

New installations, updates as well as major interventions on the configuration

of the ergo_bike medical8 may only be carried out by authorised personnel.

 Potential Equalisation

Connecting of electronic devices through V 24 / RS232 / USB / LAN, for instance,

or other interfaces of a computer, or of another independent measuring device

may only be carried out by authorised personnel and in rooms intended for

medical use according to DIN VDE (German Electrotechnology Federation)

0100 / Part 710.

Parallel connection over potential equalisation makes sure that in the case of

an eventual insulation breakdown none of the networked devices will develop

on its housing a voltage in excess of 10mV, and thus harm the patient. For this

purpose all the medical devices of daum electronic gmbh are equipped with a

connector for potential equalisation; this connector must be connected by means

of a potential equalisation cable to all the other devices in this system and to

the potential equalisation bus of the room.

Other Notes

All the diagrams and evaluations are to be considered as a suggestion from the

device. When providing a diagnosis the results must always be controlled and

evaluated by a qualified physician.

The electronic installations of the rooms where the device will be used must

absolutely comply with the requirements of directive DIN VDE (German

Electrotechnology Federation) 0100 / Part 710.

The device is not suited for operation in rooms and / or areas with explosion

risks.

All the interventions in the device, modifications to system components, extensions

to the system, as well as needed internal cleaning and repairs may only be

carried  out  by  personnel  authorised  by  daum  electronic  gmbh.

When using the device you must absolutely follow (respect) the instructions in

the user manual for the load-applying unit, particularly with respect to the safety

of the patient.

Cellular telephones may not be used in the direct vicinity of the device, otherwise

they may interfere with its functionality.

Wireless telephones may be used.

Electromagnetic interference are primarily characterised by the oscillation of the

displayed load value. If the displayed value varies uncontrollably frequently,

even when the speed is higher than 30 /min, this can be an effect of electromagnetic

compatibility (EMC).

Only cables certified by daum electronic gmbh may be used to connect the

device with other devices.

Summary of Contents for ergo_bike medical8

Page 1: ...medical Operating Instructions 0650682e 0004 0297...

Page 2: ...tting Personal Data 8 Starting Training by Selecting a Program 9 Training Programs 9 10 Setting Limit Values 10 11 Reviewing Stored Training Data 11 MP3 Player 12 Keywords Glossary 12 General Notes Im...

Page 3: ...de come into contact with others conductive parts including the grounding while getting on the ergo_bike medical8 Please do never touch device parts conducting very low safety voltage and the patient...

Page 4: ...estion from the device When providing a diagnosis the results must always be controlled and evaluated by a qualified physician The electronic installations of the rooms where the device will be used m...

Page 5: ...rs 3 Handlebars continuously adjustable over 360 4 Clamp screw for the handlebars 5 Roller casters 6 Name plate 7 Potential equalisation 8 On Off switch 9 Height adjustment for the saddle 10 Saddle co...

Page 6: ...ightening Pull and turn the lever to point to the rear risk of injury When switched on the system takes about 1 minute to load the software The colour screen displays during this time a software load...

Page 7: ...d gear The value that can be actually modified will be colour highlighted in the training view 11 Audio connector Used to connect a headphone 12 Card reader for the ergo_memo card2 Underside 13 Reset...

Page 8: ...ys Use the info key to obtain more information about the individual menu items If the requested information spans many pages you can leaf through the pages by turning the control button Press the Info...

Page 9: ...e With the RPM controlled programs the wheel speed in revolutions per minute is maintained at a level defined by the user Training Programs Pulse Rate Training Programs With the pulse rate programs th...

Page 10: ...or before setting the loading values after a few training units Dynamic Limit Values and Static Limit Values can be set Due to their natural variability dynamic limit values are continually exceeded S...

Page 11: ...ton to select Setting Turn and press control button to select Personal data Turn and press control button to Select Training Unit Turn and press control button to Select Diagram Reviewing Stored Train...

Page 12: ...rol button to start playing the selected title The system continues with playing all the music titles eventually following the selected one in the list The volume is controlled by means of the 2 dedic...

Page 13: ...should still be slightly bent at the lower pedal position Fig A as for normal biking as opposed to the racing posture Fig B Environmental Conditions Ergometers are not suited for operation in areas o...

Page 14: ...n Issue EN 1041 Informationen supplied by the manufacturer 1998 with medical devices EN 60601 1 A1 A2 Medical electrical equipment Part 1 General 1996 requirements for safety EN 60601 1 2 Medical elec...

Page 15: ...plate located on the housing with your local supply voltage prior to plugging the power cord to the power supply Please contact you dealer if the values are not the same The device is completely disco...

Page 16: ...for operating medical products and the known technical regulations and be repaired as needed Technical safety tests may only be executed by persons technicians with the proper education and knowledge...

Page 17: ...upervision duty verified II Functions ON OFF switch verified EMERGENCY STOP device verified Controls mechanical verified Controls dashboard console verified III User manual The user manual was read an...

Page 18: ...heet for the testing before placing into service and repeat testing of medical electrical devices or systems according to GEF German Electrotechnology Federation 0751 1 Next test date Remark Tester Si...

Page 19: ...ty of medical products at daum electronic gmbh The device must be tested according to GEF German Electrotechnology Federation 0751 1 before it is put back into service Failure description Spare part I...

Page 20: ...and only if you received the program in object code or executable form with such an offer in accord with Subsection b above The source code for a work means the preferred form of the work for making m...

Page 21: ...distribution limitation excluding those countries so that distribution is permitted only in or among countries not thus excluded In such case this License incorporates the limitation as if written in...

Page 22: ...f any to sign a copyright disclaimer for the program if necessary Here is a sample alter the names Yoyodyne Inc hereby disclaims all copyright interest in the program Gnomovision which makes passes at...

Page 23: ......

Page 24: ...daum electronic gmbh Flugplatzstr 100 D 90768 F rth Tel 49 0 911 97536 0 Fax 49 0 911 7537 14 www daum electronic de Stand 25042007...

Reviews: