Offset
: Initial: ±0.05% of full scale; vs.Temperature: ±25 ppm/°C; vs.Time: ±20 ppm/month
Gain Accuracy
: ±0.02% of full scale
Gain Stability
: vs. Temperature: ±25 ppm/°C; vs. Time: ±20 ppm/month
Ripple and Noise
: Readings are within the stated accuracy from 10% to 100% of the fre-
quency range in use
Filter (per channel)
: 3-pole modified Butterworth; 3 dB down at 2 Hz; 60 dB down at
25 Hz
Step-Response Settling Time (Full-Scale Output)
:
To 1% of final value: 600 msec
To 0.1% of final value: 750 msec
To 0.02% of final value: 1 sec
Auxiliary Output
: Filtered output available on mainframe wire-wrap pin
Table 1 10A40 “Type” Codes
Full-Scale
Channel
Range
Type Code
250 Hz
40
500 Hz
41
1000 Hz
42
2000 Hz
43
4000 Hz
44
8000 Hz
45
16000 Hz
46
32000 Hz
47
2
T
RANSDUCER
C
ONNECTIONS
2.a S
TANDARD
C
ABLING
The Model 10A40’s I/O CONNECTOR mates with Daytronic CONDITIONER CONNEC-
TOR No. 60322, shown in Fig. 1.5 (in Manual Section 1.E.1). Fig. 1(a) shows recom-
mended cabling for an intrinsically grounded transistor or logic-circuit driver; Fig. 1(b)
for a pulse transformer transducer with two-wire isolated windings (tachometer, tur-
bine flowmeter, etc.); and Fig. 1(c) for a “zero-velocity” (true digital output) sensor
requiring 10-V excitation. Table 2 gives standard pin assignments for the I/O Connec-
tor.
Table 2 Model 10A40 Pin Assignments
I/O Connector
Conditioner
Pin
Line
Number
Function
1
+EXCITATION (+5 V-DC)
A
–EXCITATION (-5 V-DC)
2
POWER COMMON
B
Not Committed
3
+SIGNAL
C
–SIGNAL
4,D
Not Committed
5
POWER COMMON
E
0.1-µF COUPLING
6
Not Committed
F
10-µF COUPLING
7-10
Not Committed
H-L
Not Committed
10A40.
2
T
RANSDUCER
C
ONNECTIONS
10A40
F
REQUENCY
C
ARD
Summary of Contents for System 10
Page 31: ...GETTING YOUR SYSTEM 10 ON THE AIR SB 2 FOR A SIZED MAINFRAMES...
Page 67: ...GETTING YOUR SYSTEM 10 ON THE AIR SB 2 FOR B SIZED MAINFRAMES...
Page 122: ...THIS PAGE INTENTIONALLY BLANK...
Page 134: ...1 12 THIS PAGE INTENTIONALLY BLANK...
Page 142: ...1 20 THIS PAGE INTENTIONALLY BLANK...
Page 150: ...1 28 THIS PAGE INTENTIONALLY BLANK...
Page 174: ...10A16 4C 6 THIS PAGE INTENTIONALLY BLANK 10A16 4C QUAD PLATINUM RTD CARD...
Page 186: ...10A18 4C 8 THIS PAGE INTENTIONALLY BLANK 10A18 4C QUAD PLATINUM LINEAR RTD CARD...
Page 212: ...10A40 8 THIS PAGE INTENTIONALLY BLANK 10A40 FREQUENCY CARD...
Page 224: ...10A43 4 THIS PAGE INTENTIONALLY BLANK 10A43 DWELL ANGLE CARD...
Page 234: ...10A48 6 THIS PAGE INTENTIONALLY BLANK 10A48 FLOW CARD...
Page 242: ...10A61 2 4 THIS PAGE INTENTIONALLY BLANK 10A61 2 DUAL 4 20 MA CARD...
Page 252: ...10A63 2 6 THIS PAGE INTENTIONALLY BLANK 10A63 2 DUAL DC VOLTAGE CARD...
Page 270: ...10A69 4 6 THIS PAGE INTENTIONALLY BLANK 10A69 4 QUAD AC RMS CARD...
Page 276: ...10A70 2 6 THIS PAGE INTENTIONALLY BLANK 10A70 2 DUAL DC STRAIN GAGE CARD...
Page 288: ...10A72 2C 12 THIS PAGE INTENTIONALLY BLANK 10A72 2C ENHANCED DUAL DC STRAIN GAGE CARD...
Page 320: ...10A74 4C 18 THIS PAGE INTENTIONALLY BLANK 10A74 4C QUAD DC STRAIN GAGE TRACK HOLD CARD...
Page 342: ...10A96 8 THIS PAGE INTENTIONALLY BLANK 10A96 AMPLIFIED ACCELEROMETER VIBRATION CARD...
Page 362: ...AA30 4 12 THIS PAGE INTENTIONALLY BLANK AA30 4 QUAD LVDT CARD...
Page 374: ...AA41 2 4 12 THIS PAGE INTENTIONALLY BLANK AA41 2 AA41 4 DUAL QUAD FREQUENCY CARD...
Page 390: ...AA72 2 4 16 THIS PAGE INTENTIONALLY BLANK AA72 2 AA72 4 DUAL QUAD DC STRAIN GAGE CARD...
Page 394: ...1 32 THIS PAGE INTENTIONALLY BLANK...
Page 404: ...1 42 THIS PAGE INTENTIONALLY BLANK...