325.6610.17
- 84 -
If
CALIBRATION is good, then the sensor and its electronics are ok and the problem may
be the optional gas generator (if so equipped).
• If
the
CALIBRATION failed, expose the sensor to a known gas concentration in the run
mode with the alarms disabled.
• If the measured gas concentration reads below about 50% of the expected value, check the
sensor housing openings for dust or other obstructions to gas flow. Check the sensor itself
for dust or obstructions.
• If dust is present, remove it by directing a gentle air stream close to and parallel to the
sensor face, do not blow directly onto the sensor face.
• Repeat the test with the known gas concentration.
• If the problem has not been corrected, replace the sensor.
• Determination of a faulty gas generator can only be made if the sensor passes its
CALIBRATION procedure above.
• If the sensor passes calibration, check the gas generator wiring for continuity.
• Perform a SENSOR CHECK. Within 40 seconds after pressing ENTER To start the test,
measure the voltage across the gas generator, it should not exceed 6 VDC. If the wiring
is good and the measured voltage across the gas generator exceeds 6 VDC, the gas
generator must be replaced.
Represented by:
Design improvements may be made without notice.
DEC 2018
De Nora Water Technologies
3000 Advance Lane Colmar, PA 18915
ph +1 215 997 4000 • fax +1 215 997 4062
web:
www.denora.com
mail:
info.dnwt@denora.com
®Registered Trademark. © 2018. All Rights Reserved.
Summary of Contents for Capital Controls CHLORALERT 17CA3000 Series
Page 14: ...325 6610 17 14 Figure 2 1 Receiver Outline Dimensions ...
Page 15: ... 15 325 6610 17 Figure 2 2 Sensor Transmitter Outline Dimensions ...
Page 16: ...325 6610 17 16 Figure 2 3 Interconnection Diagram ...
Page 18: ...325 6610 17 18 Figure 2 4 Sensor Wiring ...
Page 44: ...325 6610 17 44 ...