Leaf Porometer Operator’s Manual
4. Sensor Head Calibration
42
10 readings, then there is a problem with the calibration and
the following screen will appear.
The most common problem is that thermal equilibrium
hasn't been reached. Let the porometer equilibrate for a
few minutes and repeat the calibration process. If you still
can't achieve a good calibration, then contact Decagon for
help (support@decagon.com).
13) Calibration Complete
Once the calibration is complete, the new calibration num-
ber is automatically saved into the handheld unit and the
following screen is displayed. At this point, you are ready to
take measurements with your leaf porometer. Note that the
most recent saved calibration number will be applied to all
subsequent measurements until a new calibration is com-
pleted or you request that the factory default calibration
number be applied.