Leaf Porometer Operator’s Manual
8. Troubleshooting
79
Downloading data stops in the middle with an error mes-
sage saying the utility lost connection with the Leaf Porom-
eter.
SOLUTION:
A noisy serial connection can disrupt the connection
between the Utility and the Leaf Porometer. If this error
happens regularly, you can try setting your baud rate lower
or increasing the number of times a command is sent to the
Leaf Porometer. Choose the "Communications" tab in Pref-
erences to alter this.
12. PROBLEM:
No information is being displayed.
SOLUTION:
Any conductivity that is lower that 5 mmol/m
2
s
1
will not be
displayed in Automatic mode. Therefore, check to see what
mode you're in before beginning. It is advisable to use Man-
ual mode if you know that your leaf has a conductivity that
is lower than the minimum threshold for Auto mode. Also,
remember that you must press “Enter” in Automatic mode
before you begin taking a reading. You may also have gotten
moisture inside the sensor assembly or Teflon disk by
touching it with your finger. This will distort the sensor's
abilities to take accurate readings.
13. PROBLEM:
You are not getting the conductivity reading you expected.
SOLUTION: