12. Troubleshooting
50
1. PROBLEM:
WP4C won’t turn on.
SOLUTION:
•
Check to make sure your power cord is securely
attached the back of the instrument, and into the
power outlet.
•
A power surge may have caused a fuse to blow.
To
change the fuses, follow these instructions:
1. Unplug the power cord from the wall and the instru-
ment.
2. Locate the panel where the power cord plugs in. The
fuse box is on the right side of that panel. Press in on
the release tab and pull the fuse-holder out.
3. Pull the broken fuse(s) out and replace with a 2.0 Amp
250V fuse.
Caution: Do not use any other kind of fuse or you will risk damage
to your instrument and void your warranty
.
4. Replace the fuse-holder and push it into the fuse-well
until the release tab snaps in place.
5. Re-connect the power cord and turn your instrument